spaschal on DSK3GDR082PROD with CFR Pt. 36, App. A 14 CFR Ch. I (1-1-19 Edition) allowance does not eliminate the requirement of including microphone extension cables when performing the pink noise recording in section A36.3.9.5. A36.3.6.4 For analog tape recordings, the amplitude fluctuations of a 1 kHz sinusoidal signal recorded within 5 dB of the level corresponding to the calibration sound pressure level must not vary by more than Section0.5 dB throughout any reel of the type of magnetic tape used. Conformance to this requirement must be demonstrated using a device that has time-averaging properties equivalent to those of the spectrum analyzer. A36.3.6.5 For all appropriate level ranges and for steady sinusoidal electrical signals applied to the input of the measurement system, including all parts of the microphone system except the microphone, at one-thirdoctave nominal midband frequencies of 50 Hz, 1 kHz and 10 kHz, and the calibration check frequency, if it is not one of these frequencies, the level non-linearity must not exceed Section0.5 dB for a linear operating range of at least 50 dB below the upper boundary of the level range. NOTE 1: Level linearity of measurement system components may be tested according to the methods described in IEC 61265 as amended. NOTE 2: Microphone extension cables configured in the field need not be included for the level linearity determination. A36.3.6.6 On the reference level range, the level corresonding to the calibration sound pressure level must be at least 5 dB, but no more than 30 dB less than the upper boundary of the level range. A36.3.6.7 The linear operating ranges on adjacent level ranges must overlap by at least 50 dB minus the change in attenuation introduced by a change in the level range controls. NOTE: It is possible for a measurement system to have level range controls that permit attenuation changes of either 10 dB or 1 dB, for example. With 10 dB steps, the minimum overlap required would be 40 dB, and with 1 dB steps the minimum overlap would be 49 dB. A36.3.6.8 An overload indicator must be included in the recording and reproducing systems so that an overload indication will occur during an overload condition on any relevant level range. A36.3.6.9 Attenuators included in the measurement system to permit range changes must operate in known intervals of decibel steps. A36.3.7 Analysis systems. A36.3.7.1 The analysis system must conform to the specifications in sections A36.3.7.2 to A36.3.7.7 for the frequency bandwidths, channel configurations and gain settings used for analysis. A36.3.7.2 The output of the analysis system must consist of one-third octave band sound pressure levels as a function of time, obtained by processing the noise signals (preferably recorded) through an analysis system with the following characteristics: (a) A set of 24 one-third octave band filters, or their equivalent, having nominal midband frequencies from 50 Hz to 10 kHz inclusive; (b) Response and averaging properties in which, in principle, the output from any onethird octave filter band is squared, averaged and displayed or stored as time-averaged sound pressure levels; (c) The interval between successive sound pressure level samples must be 500 ms Section5 milliseconds(ms) for spectral analysis with or without slow time-weighting, as defined in section A36.3.7.4; (d) For those analysis systems that do not process the sound pressure signals during the period of time required for readout and/or resetting of the analyzer, the loss of data must not exceed a duration of 5 ms; and (e) The analysis system must operate in real time from 50 Hz through at least 12 kHz inclusive. This requirement applies to all operating channels of a multi-channel spectral analysis system. A36.3.7.3 The minimum standard for the one-third octave band analysis system is the class 2 electrical performance requirements of IEC 61260 as amended, over the range of one-third octave nominal midband frequencies from 50 Hz through 10 kHz inclusive (incorporated by reference, see Section 36.6). NOTE: IEC 61260 specifies procedures for testing of one-third octave band analysis systems for relative attenuation, antialiasing filters, real time operation, level linearity, and filter integrated response (effective bandwidth). A36.3.7.4 When slow time averaging is performed in the analyzer, the response of the one-third octave band analysis system to a sudden onset or interruption of a constant sinusoidal signal at the respective one-third octave nominal midband frequency, must be measured at sampling instants 0.5, 1, 1.5 and 2 seconds(s) after the onset and 0.5 and 1s after interruption. The rising response must be Section4 Section1 dB at 0.5s, Section1.75 Section0.75 dB at 1s, Section1 Section0.5 dB at 1.5s and Section0.5 Section0.5 dB at 2s relative to the steady-state level. The falling response must be such that the sum of the output signal levels, relative to the initial steady-state level, and the corresponding rising response reading is Section6.5 Section1 dB, at both 0.5 and 1s. At subsequent times the sum of the rising and falling responses must be Section7.5 dB or less. This equates to an exponential averaging process (slow time-weighting) with a nominal 1s time constant (i.e., 2s averaging time). A36.3.7.5 When the one-third octave band sound pressure levels are determined from 782 VerDate Sep<11>2014 12:50 Apr 30, 2019 Jkt 247046 PO 00000 Frm 00792 Fmt 8010 Sfmt 8002 Y:\SGML\247046.XXX 247046